Experiment 2: Capacitors
Objectives of the experiment
Getting to know the following components
Digital multimeter
Function generator
Oscilloscope
Breadboard
electrical-engineering learning outcome in
Preparation for the lab
in the ILIAS course
Read the documents for Experiment 2 here.
They will be made public one week before the experiment.
Function generator
Abb. 1: Function generator HM8030-5
A function generator provides a variable voltage source. In general, these signals can be generated with different waveform shapes, frequencies, and amplitudes. These values can be adjusted on real function generators. In contrast to an ideal function generator, the output current of a real system is limited. As with a real voltage source, an output impedance is specified here.
In Abbildung 1 the function generator used in the lab is shown. It has an output impedance of $50~\rm \Omega$. In the following, the settings are briefly described:
The waveform can be selected using the FUNCTION button. Pressing the button selects the next waveform. At start-up, the sine waveform (∿) is selected; the following waveforms are: triangle and square or pulse. The waveforms can be seen in the simulation below; a sawtooth signal is not possible with this function generator.
The frequency can be changed via two inputs
The potentiometer under FREQUENCY allows precise adjustment. Turning clockwise increases the frequency.
Using the buttons under the potentiometer, the frequency can be changed by one decade—i.e., by a power of ten—down (button with arrow to the left) or up (button with arrow to the right). The limits are $50~\rm mHz$ and $5~\rm MHz$.
There are also several controls for the voltage
The potentiometer
OFFSET allows precise selection of the
DC component. To enable a
DC component, press the
ON button.
At the output jack, two attenuations of $-20~\rm dB$ can be switched on. This reduces the peak-to-peak voltage range from [$0~\rm V$, $10~\rm V$] to [$0~\rm V$ , $1~\rm V$ ] or [$0~\rm V$ , $0,1~\rm V$ ].
The AMPLITUDE, i.e. the peak-to-peak voltage, can be finely adjusted using a potentiometer.
Oscilloscope used
Even before the digital multimeter, the oscilloscope is the most important measuring instrument in electrical engineering and electronics. It makes it possible to display a voltage waveform u(t) over time t, observe it in “real time,” and measure it. In many experiments and analyses, it is a central component because it can make electrical processes visible. In addition to quantitative statements (how high is the voltage and when?), it is also helpful for qualitative results (for example: Is there a fault in the circuit?).
Abb. 2: Display and control panel of the oscilloscope DS1052E
Good knowledge of the oscilloscope is necessary for this experiment. In Abbildung 2 you can see the control panel of the DS1052E used, which is briefly described here.
Please use the buttons marked with „ “ to learn more about the individual functions
General operation
On/off switch (1.)
Display area (2.)
Selection of operating modes (3.)
Menu selection buttons (4.)
The oscilloscope can be turned on/off using the button on the top (1.)
In the display area (2.) information is shown in the user interface after switching on, which is explained below
At the top right (3.) two operating modes can be selected: single acquisition after a trigger signal (RUN/STOP) and automatic selection of various settings (AUTO). In addition, there are various menu selection buttons there, e.g. for the cursor menu (3.).
In the menus, displayed functions can be selected via five selection buttons (4.). The menus are not discussed further below.
Signal display and input
The largest area is the display window (5.). The display window shows 10 horizontal and 8 vertical divisions (eng. divisions, abbreviation Div.). As a rule, the voltage-time waveform is shown here.
In the figure, two waveforms are displayed simultaneously: channel 1 in yellow and channel 2 in mint. These can be switched on/off using the channel selection (6. top two buttons).
The signals are fed in via the input jacks (7.) (CH 1 : channel 1, CH 2 : channel 2)
For the signal, you can choose whether the
DC component is suppressed (
AC coupling) or not (
DC coupling). The
channel coupling is shown in the display (8.) as a line (
= for
DC coupling) or a tilde (
~ for
AC coupling)
Scaling
vertical controls (9.)
horizontal controls (6.)
voltage scale (8.)
time base (10.)
signal mean values (11.)
The voltage scale can be varied using the vertical controls (9.): POSITION shifts a waveform up/down. SCALE enlarges/reduces the vertical axis (= voltage axis).
The same can be done for the time axis using the horizontal controls (6.): it can also be shifted with POSITION and enlarged/reduced with SCALE.
The current scaling can be read from the display area: In the image, a voltage scale of 2.00 V/Div for both signal waveforms (8.) and a time base of 50.00 us/Div (10.) are shown.
The signal mean values are drawn to the left of the waveform (11.) as a small yellow arrow (channel 1) and a mint-colored arrow (channel 2). If a signal is above or below the displayed area due to incorrect scaling, the corresponding arrow is also shown at the top or bottom edge.
Trigger
As a rule, acquisition should be triggered when a certain threshold is exceeded or fallen below. This threshold is called the trigger threshold.
The trigger threshold, trigger source (CH1 or CH2), and triggering on an exceedance (rising edge ͟ ↑͞ ) or falling below (falling edge ͞ ↓͟ ) can be seen in the display area (12.). The threshold is additionally marked to the left of the waveform (11.).
The knob (13.) can be used to shift the threshold (trigger level). Pressing 50% sets the trigger to the center.
The acquisition status (14.) is located at the top left of the display area. STOP means a still image is shown. When running, T'D for triggered is shown here.
The oscilloscope records a longer time period than the one displayed. The displayed position in memory (15.) is shown at the top of the display area. The trigger point is also drawn there. How many (nano-, micro-, or milli-)seconds the trigger point is currently shifted from the displayed center is shown at the bottom right (16.).
On the left you will find a nice introductory video. Note that the specific operation often depends on the manufacturer and model. However, the concepts are the same for all devices.
Virtual oscilloscopes
A virtual oscilloscope can be found on the pages of Aachen University. Try there to “oscilloscope” various function generator settings, e.g.:
$200~\rm Hz$, sine, offset $1~\rm V$, amplitude $5~\rm V$
$200~\rm Hz$, square, offset -$1~\rm V$, amplitude $3~\rm V$
What happens if the trigger level is set too high?
further virtual oscilloscopes
Hochschule Aalen also provides a virtual oscilloscope for practice.
Another virtual oscilloscope is offered by TU Berlin. However, due to the Java implementation it only works in a few browsers (e.g. Internet Explorer). In addition, the following must be done:
Preparation for the oral short test
For this experiment you should
be able to apply and explain the following concepts:
periodic signal
characteristic quantities in the signal-time waveform
peak value (amplitude)
peak-to-peak value
arithmetic mean value
RMS value (quadratic mean)
rectified mean value
period duration
frequency
creating the time waveform of a periodic signal
graphical determination of the above quantities from the time waveform of a signal
capacitance $C$
behavior of current $i$ and voltage $u$ across $R$ and $C$ with a rectangular input voltage for $t=[0; \infty]$
$i_{C,stat}$ and $u_{C,\rm stat}$ in steady state
time constant $\tau$
determination of $R$, $C$, $\tau$, $i_{C,\rm stat}$ and $u_{C,\rm stat}$ from the time waveform